Hoefer HE33 User Manual Page 13

  • Download
  • Add to my manuals
  • Print
  • Page
    / 26
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 12
p5
Cast the  gel
Install the running tray
Firmly grasp the casting tray with one hand. With
the other hand, place one end of the running tray
against the foam pad at the bottom edge, press the
tray against the pad, and then lower it to rest on the
bottom of the casting tray, seating the other end of
the tray against the opposite foam pad.
Prepare the comb(s)
Fit the two slots in the comb between the (loosened)
thumb screw heads and the comb back. Tighten the
screws until the comb is just supported. Seat the
comb assembly on the rim of the casting tray and
adjust the bottom of the comb so that it is about
1.0 mm from the running tray. Tighten the screws
to secure the comb. To run twice as many samples,
prepare two combs.
Remove the comb assembly. Place the casting
assembly on a leveling surface and level, using the
spirit level on the running tray as a guide.
Pour the agarose solution (cooled to 50 °C) into the
casting tray. Orient the comb assembly so that the
comb faces the nearest foam pad and seat it on the
tray rim. Check that the comb is vertical to prevent
well shape distortions. To run twice as many samples,
place the second comb assembly in the center of
the tray. Allow a minimum of 30 minutes for the gel
to set.